Wire electrometers are less sensitive by about two orders of magnitude
than those with quadrants, but they have the advantage of bringing together
great speed, great simplicity of use and good stability. For these reasons
they were favoured over quadrant electrometers in all cases in which sensitivity
as great as that of quadrant electrometers was not required.
In the Lutz-Edelmann electrometer the auxiliary field is created
by a pair of steel electrodes with a triangular cross section, arranged
vertically and well isolated, with two edges, lengthened by a slight curvature
of the lateral faces, placed side by side (blades).
Between the two blades there is a very thin platinum wire which is held
taut by an insulating quartz reed and an elastic string, it too made of
quartz, closed in a loop.
The blades and the wire are protected and insulated by a metal case which
is earthed. In the symmetrical connection, which is the most widely employed,
the two blades are maintained at equal and opposite potentials of about
one hundred volts by connecting them to a battery, the central point of
which is earthed. By moving the blades closer together or farther apart,
it is possible to create a more or less intense field between them....continues
|
 |
 |